Laakso O, Pennanen T, Himberg K, Kuitunen T, Himberg JJ. J. Forensic Sci. 2004; 49(5): 1113-1116.

Affiliation

Hyvinkää Hospital, Department of Anaesthesia and Intensive Care Medicine, Hyvinkää, Finland.

Copyright

(Copyright © 2004, American Society for Testing and Materials, Publisher John Wiley and Sons)

DOI

unavailable

PMID

15461120

Abstract

The Drager 7110 MK III FIN Evidential breath analyzer is classified as a quantitative analyzer capable to provide sufficient evidence for establishing legal intoxication. The purpose of this study was to evaluate ethanol specificity of this instrument in the presence of other solvents. Effects of eight possible interfering compounds on ethanol analysis were determined in a procedure simulating a human breathing. Most of the compounds studied had either a negligible effect on ethanol analysis (acetone, methyl ethyl ketone, and methyl isobutyl ketone) or were detected in very low concentrations before influencing ethanol readings (methanol, ethyl acetate, and diethyl ether). However, 1-propanol and 2-propanol increased the ethanol readings significantly. Thus, Dräger ethanol readings should be interpreted carefully in the presence of propanol.