Caplan YH, Yohman DT, Schaefer JA. J. Forensic Sci. 1985; 30(4): 1058-1063.

Copyright

(Copyright © 1985, American Society for Testing and Materials, Publisher John Wiley and Sons)

DOI

unavailable

PMID

4067535

Abstract

Ninety Breathalyzer instruments (Model 1000) and twenty instruments (Models 900, 900A) were studied using a protocol described by the Department of Transportation's "Standard for Devices to Measure Breath Alcohol." Although the mean of each of three concentrations tested (0.05, 0.10, and 0.15 g/210 L) compared favorably in both series, the standard deviation was consistently higher for the Model 1000 instruments. The Model 1000 instruments also produced a significant number of test results which exceeded the normally expected scientific deviation.