SAFETYLIT WEEKLY UPDATE

We compile citations and summaries of about 400 new articles every week.
RSS Feed

HELP: Tutorials | FAQ
CONTACT US: Contact info

Search Results

Journal Article

Citation

Goldberger BA, Caplan YH. J. Forensic Sci. 1987; 32(1): 48-54.

Copyright

(Copyright © 1987, American Society for Testing and Materials, Publisher John Wiley and Sons)

DOI

unavailable

PMID

3819688

Abstract

In vitro accuracy and precision studies were conducted using silica gel, magnesium perchlorate, and indium encapsulation breath collection tubes in conjunction with three infrared breath ethanol analyzers (BAC Verifier, Intoxilyzer 5000, and Intoximeter 3000), the Breathalyzer 900A, and the GC Mark IV. Statistical analyses revealed good accuracy and precision and correlation between direct and delayed vapor ethanol analyses for each combination of instruments and collection devices (range = 0.000 to 0.250 g/210 L, N = 42/instrument, r greater than 0.99). Delayed vapor ethanol analysis utilizing each instrument and collection device combination appears to predict satisfactorily original vapor ethanol concentrations.


Language: en

NEW SEARCH


All SafetyLit records are available for automatic download to Zotero & Mendeley
Print