SAFETYLIT WEEKLY UPDATE

We compile citations and summaries of about 400 new articles every week.
RSS Feed

HELP: Tutorials | FAQ
CONTACT US: Contact info

Search Results

Journal Article

Citation

Baugh LD, Allen EE, Liu RH, Langner JG, Fentress JC, Chadha SC, Cook LF, Walia AS. J. Forensic Sci. 1991; 36(1): 79-85.

Affiliation

Department of Criminal Justice, University of Alabama, Birmingham.

Copyright

(Copyright © 1991, American Society for Testing and Materials, Publisher John Wiley and Sons)

DOI

unavailable

PMID

2007883

Abstract

Immunoassay kits for urine cocaine (and metabolite) screening, obtained from two commercial sources, were examined for correlation of their results, expressed in terms of equivalent benzoylecgonine concentration, with the gas chromatography/mass spectrometry (GC/MS) concentration of benzoylecgonine. The correlation coefficients obtained, based on 62 (out of a total sample population of 3295) highly relevant samples, were 0.467 and 0.766 for Abuscreen (ARIA) and TDx (TDX), respectively. The preliminary screen cutoff values, which correspond to 150 ng/mL benzoylecgonine (as determined by GC/MS), were calculated based on the resulting regression equations and found to be 380 and 190 ng/mL for ARIA and TDX, respectively. With these cutoff values, ARIA generates 5 false negatives and 16 unconfirmed presumptive positives, while TDX results in 3 false negatives and 6 unconfirmed presumptive positives.


Language: en

NEW SEARCH


All SafetyLit records are available for automatic download to Zotero & Mendeley
Print