@article{ref1,
title="Field performance of the Intoxilyzer 5000: a comparison of blood- and breath-alcohol results in Wisconsin drivers",
journal="Journal of forensic sciences",
year="1990",
author="Harding, P. M. and Laessig, R. H. and Field, P. H.",
volume="35",
number="5",
pages="1022-1028",
abstract="Intoxilyzer 5000 and blood-alcohol results from drivers arrested for operating a motor vehicle while intoxicated and for related offenses were compared during a two-year period. Three hundred and ninety-five pairs of results were studied. The breath- and blood-alcohol specimens in this study were collected within 1 h of each other. The mean blood-alcohol concentration obtained was 0.180 g/dL, with a range from zero to 0.338 g/dL. By comparison, the mean Intoxilyzer 5000 result was 0.16 g/210 L with a range from zero to 0.32 g/210 L. Compared with the blood-alcohol result, Intoxilyzer 5000 results were lower by more than 0.01 g/210 L 67% of the time, within 0.01 g/210 L 31% of the time, and higher by more than 0.01 g/210 L 2% of the time.<p /> <p>Language: en</p>",
language="en",
issn="0022-1198",
doi="",
url="http://dx.doi.org/"
}