@article{ref1,
title="Toluene and the intoxilyzer 5000: no response to concentrations found on human breath",
journal="Proceedings International Council on Alcohol, Drugs and Traffic Safety Conference",
year="2000",
author="Glover, P.l.",
volume="2000",
number="",
pages="-p.",
abstract="The response of the Intoxilyzer 5000 (series 66, 3 filter) to known amounts of ethanol vapor, toluene vapor and a mixture of ethanol and toluene vapors was determined. The ethanol concentration approximates a Breath Alcohol Concentration (BrAC) of 0.08 gm/210 L. The toluene concentration used was approximately 5 times that reported to be found on the breath of test subjects 30 minutes post exposure. There was no response by the Intoxilyzer to the toluene alone and there was no added response to the ethanol/toluene mixture response over that of ethanol alone.<p />",
language="",
issn="",
doi="",
url="http://dx.doi.org/"
}