@article{ref1,
title="Whole-body vibration in the skeleton: development of a resonance-based testing device",
journal="Annals of biomedical engineering",
year="1997",
author="Fritton, J. C. and Rubin, C. T. and Qin, Y. X. and McLeod, K. J.",
volume="25",
number="5",
pages="831-839",
abstract="Whole-body vibration (WBV) has been demonstrated to have a strong influence on physiological systems, ranging from severely destructive to potentially beneficial. Unfortunately, the study of WBV in a controlled manner is commonly constrained by space and budgetary factors, particularly where vibration in the low frequency range is considered. In the work presented here, a small, low-cost device for performing WBV of the human skeleton is developed to assist in studies of vertical acceleration in a clinical setting. The device design consists of a spring-supported plate driven by an 18 N peak-force electromagnetic actuator, and the associated driving and monitoring electronics. Animal and human lumped-mass models have been coupled with a model of the loading device to seek a resonance response in the vicinity of 30 Hz. This approach minimizes the loading requirements of such a device, and thus a major component of the cost, yet can provide peak accelerations of 0.15 g at a frequency of 30 Hz in a small, lightweight package capable of use in a clinical or laboratory setting.<p /><p>Language: en</p>",
language="en",
issn="0090-6964",
doi="",
url="http://dx.doi.org/"
}