TY - JOUR PY - 1985// TI - Portable emergency ventilators. 2. Sensitivity to environment JO - Acta anaesthesiologica Scandinavica A1 - Dahlgren, B. E. A1 - Nilsson, H. G. A1 - Peters, Bjorn A1 - Skedevik, C. SP - 753 EP - 757 VL - 29 IS - 8 N2 - Three compact ventilators (Oxylog, Maxaman and Pneupac 2), commercially available in Sweden, were subjected to tests concerning sensitivity to environmental stress. The ventilators functioned very well at temperatures above 0 degree C. The Oxylog apparatus was very sensitive to subzero temperatures. It also had a spectrum of resonance frequencies when subjected to vibration testing that restricts its use outside the sheltered hospital environment. The Maxaman and Pneupac 2 in the modified version withstood the tests in a reliable way. The shock tests performed had no negative effects on any of the ventilators or their functions.
Language: en
LA - en SN - 0001-5172 UR - http://dx.doi.org/ ID - ref1 ER -