TY - JOUR PY - 1991// TI - Precision of blood alcohol determination using gc and adh methods JO - Blutalkohol A1 - Schmitt, G. A1 - Schmidt, G. A1 - Aderjan, R. SP - 325 EP - 328 VL - 28 IS - 5 N2 - For the precision testing of the gas-chromatographic and the ADH-enzymatic method respectively we used commercially available control sera. During one year of testing (1990) they were daily and routinely used together with the specimen analysed. With the automatized GC-method the precision required according to the jurisdiction of the Bundesgerichtshof was better than with the manual ADH-method. The precision of the ADH-method should be improved using an autoanalyzer. Systematical errors between the two different methods (GC and ADH) can be minimized by a double GC-determination. Our investigation also aimed at a possible lower threatening threshold of 0.5% (BAK) as a consequence of the german "Wiedervereinigung". The precision for this concentration is 0.02% for the GC- and the ADH-method.
Language: de
LA - de SN - 0006-5250 UR - http://dx.doi.org/ ID - ref1 ER -