TY - JOUR PY - 2016// TI - Regulation of x-ray security scanners in Michigan JO - Health physics A1 - Parry, Donald E. SP - S9 EP - S16 VL - 110 IS - 2 Suppl 1 N2 - In January of 2013 the Transportation Security Administration (TSA) ordered the removal of x-ray security scanners from airports by June of 2013. Since that time several of these scanners have been purchased at a reduced cost by various state and county governments for use in screening individuals entering or leaving their facilities. To address this issue the Radiation Safety Section of the State of Michigan drafted a set of registration conditions for facilities to follow when using these security scanners. Inspection procedures and measurement protocols were developed to estimate the dose to screened individuals. Inspections were performed on nine of the 16 registered backscatter scanners in the state and the one transmission scanner. The average estimated effective dose to screened individuals was ∼11 nSv for a two view scan from a backscatter system. The effective dose was 0.446 μSv, 0.330 μSv, and 0.150 μSv for a transmission system operated in the high, medium, and low dose modes, respectively. The limit suggested in the new registration condition is 0.25 μSv for a general use system and 10 μSv for a limited use system.
Language: en
LA - en SN - 0017-9078 UR - http://dx.doi.org/10.1097/HP.0000000000000400 ID - ref1 ER -