Mehdirad A, Love C, Nelson S, Schaal S, Collins J, Huffman K. Pacing Clin. Electrophysiol. 1997; 20(7): 1885-1886.

Affiliation

Department of Medicine, Ohio State University Medical Center, Ohio State University, Columbus 43210, USA.

Copyright

(Copyright © 1997, John Wiley and Sons)

DOI

unavailable

PMID

9249849

Abstract

A patient with an ICD accidentally grasped a power line and was electrocuted. He was unable to release the cable during electrocution though he remained conscious. After receiving a shock from his ICD, the powerline was released. ICD interrogation revealed inappropriate detection of alternating current and delivery of a shock.