SAFETYLIT WEEKLY UPDATE

We compile citations and summaries of about 400 new articles every week.
RSS Feed

HELP: Tutorials | FAQ
CONTACT US: Contact info

Search Results

Journal Article

Citation

O'Connor SM, Baweja HS, Goble DJ. J. Biomech. 2016; 49(16): 4142-4145.

Affiliation

School of Exercise and Nutritional Sciences, San Diego State University, 5500 Camanile Drive, San Diego, CA 92182, United States. Electronic address: dgoble@mail.sdsu.edu.

Copyright

(Copyright © 2016, Elsevier Publishing)

DOI

10.1016/j.jbiomech.2016.10.020

PMID

27789036

Abstract

The BTrackS Balance Plate (BBP) is a low-cost force plate that provides objective balance assessment and true portability for the user. Given that this technology is relatively new, the purpose of the present study was to provide the first center of pressure (COP) validation of the BBP. Two BBP devices (one new and one used) were compared with a laboratory-grade force plate (LFP) during simultaneous collection of COP that was induced by an inverted pendulum device with human-like sway characteristics. The results of this study showed almost perfect agreement between the BBP devices and the LFP (ICC>0.999), as well as a high degree of BBP accuracy (<1% error magnitude) and precision (<0.2mm regression residuals). These results suggest the BBP can serve as an effective, low-cost solution for objective balance testing in the laboratory or clinic.

Copyright © 2016 Elsevier Ltd. All rights reserved.


Language: en

NEW SEARCH


All SafetyLit records are available for automatic download to Zotero & Mendeley
Print