Abstract

Among the various incidents in industrial radiography, inadvertent handling of sources by hands is one of the most frequent incidents in which some parts of the hands may be locally exposed to high doses. An accurate assessment of extremity dose assists medical doctors in selecting appropriate treatments, preventing the injury expansion in the region. In this study, a phantom was designed to simulate a fisted hand of a radiographer when the worker holds a radioactive source in their hands. The local doses were measured using implanted TLDs in the phantom at different distances from a source. Furthermore, skin dose distribution was measured by Gaf-chromic films in the palm region of the phantom. The reliability of the measurements has been studied via analytical as well as Monte-Carlo simulation methods. The results showed that the new phantom design can be used reliably in extremity dose assessments, particularly at the points next to the source.


Language: en

Keywords

Dosimetry; Gaf-chromic; Hand; Iridium-192; Phantom; TLD