SAFETYLIT WEEKLY UPDATE

We compile citations and summaries of about 400 new articles every week.
RSS Feed

HELP: Tutorials | FAQ
CONTACT US: Contact info

Search Results

Journal Article

Citation

Bolton ML, Zheng X, Li M, Edworthy JR, Boyd AD. Hum. Factors 2019; ePub(ePub): 18720819862911.

Affiliation

The University of Illinois at Chicago, USA.

Copyright

(Copyright © 2019, Human Factors and Ergonomics Society, Publisher SAGE Publishing)

DOI

10.1177/0018720819862911

PMID

31411902

Abstract

OBJECTIVE: This research investigated whether the psychoacoustics of simultaneous masking, which are integral to a model-checking-based method, previously developed for detecting perceivability problems in alarm configurations, could predict when IEC 60601-1-8-compliant medical alarm sounds are audible.

BACKGROUND: The tonal nature of sounds prescribed by IEC 60601-1-8 makes them potentially susceptible to simultaneous masking: where concurrent sounds render one or more inaudible due to human sensory limitations. No work has experimentally assessed whether the psychoacoustics of simultaneous masking accurately predict IEC 60601-1-8 alarm perceivability.

METHOD: In two signal detection experiments, 28 nursing students judged whether alarm sounds were present in collections of concurrently sounding standard-compliant tones. The first experiment used alarm sounds with single-frequency (primary harmonic) tones. The second experiment's sounds included the additional, standard-required frequencies (often called subharmonics). T tests compared miss, false alarm, sensitivity, and bias measures between masking and nonmasking conditions and between the two experiments.

RESULTS: Miss rates were significantly higher and sensitivity was significantly lower for the masking condition than for the nonmasking one. There were no significant differences between the measures of the two experiments.

CONCLUSION: These results validate the predictions of the psychoacoustics of simultaneous masking for medical alarms and the masking detection capabilities of our method that relies on them. The results also show that masking of an alarm's primary harmonic is sufficient to make an alarm sound indistinguishable. APPLICATION: Findings have profound implications for medical alarm design, the international standard, and masking detection methods.


Language: en

Keywords

audition; medical devices and technologies; patient safety; psychophysical methods; signal detection theory