SAFETYLIT WEEKLY UPDATE

We compile citations and summaries of about 400 new articles every week.
RSS Feed

HELP: Tutorials | FAQ
CONTACT US: Contact info

Search Results

Journal Article

Citation

Shaw A, Fisher J. J. Med. Eng. Technol. 1979; 3(6): 301-302.

Copyright

(Copyright © 1979, Informa - Taylor and Francis Group)

DOI

unavailable

PMID

526521

Abstract

Investigation of a series of falls from hospital beds showed that the cause was the poor design of the particular mattresses. Interior sprung mattresses of this construction are unstable at their edges and are particularly hazardous when used with the modern, rigid base hospital bed.